Fluke 725 ex [15/54] For entity parameters refer to fluke 725ex ccd for use in ex hazardous areas

Fluke 725 ex [15/54] For entity parameters refer to fluke 725ex ccd for use in ex hazardous areas
Multifunction Process Calibrator
Specifications
7
Temperature, Thermocouples
Type Range Measure and Source Accuracies
J
-200 to 0 °C
0 to 1200 °C
1.0 °C
0.7 °C
K
-200 to 0 °C
0 to 1370 °C
1.2 °C
0.8 °C
T
-200 to 0 °C
0 to 400 °C
1.2 °C
0.8 °C
E
-200 to 0 °C
0 to 950 °C
0.9 °C
0.7 °C
R
-20 to 0 °C
0 to 500 °C
500 to 1750 °C
2.5 °C
1.8 °C
1.4 °C
S
-20 to 0 °C
0 to 500 °C
500 to 1750 °C
2.5 °C
1.8 °C
1.5 °C
B
600 to 800 °C
800 to 1000 °C
1000 to 1800 °C
2.2 °C
1.8 °C
1.4 °C
L
-200 to 0 °C
0 to 900 °C
0.85 °C
0.7 °C
U
-200 to 0 °C
0 to 400 °C
1.1 °C
0.75 °C
N
-200 to 0 °C
0 to 1300 °C
1.5 °C
0.9 °C
XK
-200 to 100 °C
-100 to 800 °C
0.5 °C
0.6 °C
BP
0 to 800 °C
800 to 2500 °C
1.2 °C
2.5 °C
Resolution:
0.1 °C, 0.1 °F
Loop Power Supply
Voltage: 12 V
Maximum current: 24 mA
Short circuit protected
Entity Parameters
For Entity Parameters, Refer to Fluke 725Ex CCD for use in Ex-hazardous areas.
RTD Excitation (simulation)
Allowable Excitation by RTD type
Ni 120 0.15 to 3.0 mA
Pt 100-385 0.15 to 3.0 mA
Pt 100-3926 0.15 to 3.0 mA
Pt 100-3916 0.15 to 3.0 mA
Pt 200-385 0.05 to 0.80 mA
Pt 500-385 0.05 to 0.80 mA
Pt 1000-385 0.05 to 0.40 mA

Содержание