Fluke 725 ex [21/54] Equipment and software required to perform the procedures in this manual are identified in table 3 if the recommended equipment model is not available other equipment can be substituted if it meets the specifications indicated

Fluke 725 ex [21/54] Equipment and software required to perform the procedures in this manual are identified in table 3 if the recommended equipment model is not available other equipment can be substituted if it meets the specifications indicated
Multifunction Process Calibrator
Required Equipment
11
Required Equipment
Equipment and software required to perform the procedures in this manual are identified
in Table 3. If the recommended equipment model is not available, other equipment can be
substituted if it meets the specifications indicated.
WWarning
To avoid safety hazards and equipment damage during the calibration
procedures, use the specified calibration equipment listed in Table 3.
Using unspecified equipment can give unreliable results and pose
safety hazards.
Table 3. Required Equipment and Software
Equipment Minimum Specifications Recommended Model
Multi-Product Calibrator DC voltage: 0 V to 30 V
Accuracy: ±0.005 %
DC current: 0 mA to 24 mA
Accuracy: ±0.005 %
Temperature: Type J thermocouple
90 day accuracy: ±0.2 °C
Resistance accuracy: ±0.006 %
Frequency accuracy: ±0.01 %
Fluke Model 5520A Multi-Product
Calibrator only
MET/CAL
®
Metrology
Software (see MET/CAL
Installation and Upgrade
Guide for minimum
hardware requirements)
Version 6.0 or later Contact Fluke for the latest version.
Digital Multimeter DC voltage: 0 V to 24 V
Accuracy: ±0.0013 %
DC voltage: 0 mA to 24 mA
Accuracy: ±0.005 %
HP 3458A onlyno substitute
Pressure Module
In working condition; for establishing
communication only
Fluke 700PXX Series
Test Leads Two sets of stackable banana test
leads
2 red leads: Fluke PN 105809
2 black leads: Fluke PN 105806
Thermocouple test lead 2 type J male miniplugs with ~12”
(30.5 cm) of 20-gauge type J
thermocouple wire
Fluke Model 80CJ-M male miniplugs
(package of two)
20-gauge type-J thermocouple wire is
available through an electronic supply
outlet
Personal computer Windows
®
3.1 or later with terminal
communication software
486 (or later) IBM-compatible
PC interface cable
Fluke 700SC serial interface cable
assembly (LEMO to DB-9 female)
Fluke PN 667425
Calibration cable Fluke 724/725Ex Calibration Cable
700SC Serial Cable
Fluke PN 1556747
Fluke PN 667425