Fluke 725 ex [43/54] Cal ma output

Fluke 725 ex [43/54] Cal ma output
Multifunction Process Calibrator
Calibration Adjustment Procedures
33
Cal mA Output
Connect the UUT as shown in Figure 16.
aal12f.eps
Figure 16. mA Output Calibration Connections
From the Calibration Menu, enter the cal step for Cal mA Output. The PC displays:
Zero into DAC. Enter mA displayed :
Set the HP 3458 to read DC current. Be sure the NPLC on the HP 3458 is slow enough to
get stable readings to four decimal places. When the reading on the HP 3458 has
stabilized, enter the value in milliamps on the PC, then press Enter. It is only necessary to
enter four places past the decimal point. It is not necessary to enter the units (mA). Wait
for the following prompt on the PC:
Max value into DAC. Enter mA value displayed :
Enter the current reading from the HP 3458 in the PC, then press Enter. After a short
while, the following calibration constants and new prompt are displayed on the PC:
-1.628906 = ( 24.735999 - 0.076400 )/ 65279.00
mA per count = -1.628906
- press space bar to continue
Press the space bar to return to the Calibration Menu.