Automatic Pressure Calibrators
Safety
3
Only assemble and operate high-pressure systems if you know the
correct safety procedures. High-pressure liquids and gases are
hazardous and the energy from them can be released without
warning.
Remove the input signals before you clean the Product.
Use only specified replacement parts.
Have an approved technician repair the Product.
Do not disassemble or crush battery cells and battery packs.
Batteries contain hazardous chemicals that can cause
burns or explode. If exposure to chemicals occurs, clean
with water and get medical aid.
Do not put battery cells and battery packs near heat or fire. Do not
put in sunlight.
Use only Fluke approved power adapters to charge the battery.
Disconnect the battery charger and move the Product or battery to
a cool, non-flammable location if the rechargeable battery
becomes hot (>50 °C) during the charge period.
Replace the rechargeable battery after 5 years of moderate use or
2 years of heavy use. Moderate use is defined as recharged twice a
week. Heavy use is defined as discharged to cutoff and recharged
daily.
Do not disassemble the battery.
Do not short the battery terminals together.
Pressure sensors can be damaged and/or personnel injury
can occur due to improper application of pressure. The
Product shows “OL” when the pressure exceeds 110 % of
the nominal range of the sensor. When “OL” is shown on
any pressure, the pressure should be reduced or vented
immediately to prevent Product damage or possible
personnel injury. Push to zero the pressure sensor
when vented to atmospheric pressure.
Do not point the vent port toward the operator during
venting.