Fluke 729 [9/46] Specifications

Fluke 729 [9/46] Specifications
Automatic Pressure Calibrators
Specifications
5
Specifications
Pressure Specification
1-year Specification ................................... 0.02 % of full scale
Control Specification ................................. 0.005 % full scale minimum
Temperature Compensation ...................... 15 °C to 35 °C (59 °F to 95 °F) to
rated accuracy
Note: For temperatures from -10 °C to
+15 °C and 35 °C to 50 °C, add
0.04 % of full scale
Electrical Specification
All specifications are valid to 110 % of range, except 24 mA source and simulate
which are valid to 100 % of range.
Ranges
mA ............................................................. 0 mA to 24 mA
Volts .......................................................... 0 V dc to 30 V dc
Resolution
mA Ranges ................................................ 1 μA
Voltage Range ........................................... 1 mV
Accuracy .................................................... 0.01 % ±2 LSD all ranges (at 23 °C ±5 °C)
Temperature Compensation ...................... 20 ppm of full scale /°C from -10 °C to
+18 °C and 28 °C to 50 °C
Loop Compliance Voltage ......................... 24 V dc @ 20 mA
mA Simulate External
Voltage Requirement ................................. 12 V dc to 30 V dc
Temperature Measurement
Only/100 Ω Pt(385) RTD ........................... -50 °C to +150 °C (-58 °F to +302 °F)
Temperature Resolution ............................ 0.01 °C (0.01 °F)
Temperature Accuracy .............................. ±0.1 °C (0.2 °F) (±0.25 °C ±0.45 °F)
combined uncertainty when using 720
RTD probe (optional accessory)
Drive Capability ......................................... 1200 Ω without HART resistor, 950 Ω
with internal HART resistor