Warning messages displayed on the DPM measurement badges
Warning Message Cause Suggestion
Incorrect Input Trigger level Not able to trigger on the
specified Input Trigger Level
as set in the DPM Turn On/
Turn Off measurement.
Adjust the Input Trigger Level to be within the range of the input
source waveform.
Output<x> Voltage Level
reached before input trigger.
Turn On Time value is
negative for the given
configuration in the DPM Turn
On/Turn Off measurement.
Adjust the Output Voltage Level given for Rail<x> or Adjust the
Input Trigger Level for a positive Turn On Time.
Incorrect Output<x> Voltage
level.
Not able to find the specified
Output Voltage level for
Rail<x> output source as set in
the DPM Turn On/Turn Off
measurement.
Adjust the Output Voltage Level given for Rail<x> such that
voltage is within the span of the Rail<x> source waveform of
the measurement.