Fluke 725 [30/80] 726 only

Fluke 726 [30/80] 726 only
724/725/726
Calibration Manual
22
Table 5. Serial Command List-726 and 725 V3.00 and Later (cont.)
Command Response/Actions
Command
Arguments
Comment Ch 1 Ch 2
CPRT_MAX_T?
(726 only)
Returns {number} CEL.
X
CPRT_COEFA
(726 only)
1 argument. Sets the
custom CPRT
Coefficient A.
X
CPRT_COEFA?
(726 only)
Returns the custom CPRT
Coefficient A.
X
CPRT_COEFB
(726 only)
1 argument. Sets the
custom CPRT
Coefficient B.
X
CPRT_COEFB?
(726 only)
Returns the custom CPRT
Coefficient B.
X
CPRT_COEFC
(726 only)
1 argument. Sets the
custom CPRT
Coefficient C.
X
CPRT_COEFC?
(726 only)
Returns the custom CPRT
Coefficient C.
X
RTD_WIRE 1 argument, 2W, 3W or
4W. Sets RTD read
wire.
X
RTD_WIRE? Returns RTD read wire. Verify
connection
setting
X
TEMP_UNIT 1 argument. Sets
temperature units, CEL
or FAR
CEL: Celsius
FAR: Fahrenheit
X
TEMP_UNIT? Returns temperature units, CEL or
FAR
X
CUST1_ALIAS
(726 only)
1 argument, sets
screen name for
CUST1 RTD.
X
CUST1_ALIAS?
(726 only)
Returns screen name for CUST1
RTD
Verify RTD 1
alias
X
CUST2_ALIAS
(726 only)
1 argument, sets
screen name for
CUST2 RTD.
X
CUST2_ALIAS?
(726 only)
Returns screen name for CUST2
RTD.
Verify RTD 2
alias
X
CUST3_ALIAS1
(726 only)
1 argument, sets
screen name for
CUST3 RTD.
X

Содержание