Calibration Manual
32
Lower Display Frequency Source Test (725 and 726)
1. Press on the UUT until SOURCE appears on the lower display.
2. Press on the UUT (for 726) until Hz appears on the lower display.
3. Configure the Fluke 8508A to measure frequency; then make the connections shown
in Figure 7.
4. Use the arrow keys on the UUT to set the UUT output 10 kHz at 5 V and verify that
the Fluke 8508A readings are within the limits 9.975 Hz to 10.025 kHz.
5. Press on the UUT to disable the sourcing function.
aal08f.eps
Figure 7. Lower Display Frequency Source Test Connections