Fluke 725 [75/80] Calibrating thermocouple mv output

Fluke 726 [75/80] Calibrating thermocouple mv output
Calibrators
Calibration Adjustment
67
Calibrating Thermocouple mV Output
Connect the UUT as shown in Figure 33.
aal08f.eps
Figure 33. Thermocouple mV Output Calibration Connections
After making the proper connections, type “10” for 726/725 or “8” for 724, and then press
the Enter key on the PC keyboard. Input directions are displayed at the bottom of the
calibration menu:
Enter Selection:
First Calibration Point. Enter mV displayed:
Second Calibration Point. Enter mV displayed:
Use the 8508A to measure the two calibration points. Enter each of the two calibration
points as prompted.