Fluke 725 [42/80] Lower display 3 wire rtd measurement

Fluke 726 [42/80] Lower display 3 wire rtd measurement
724/725/726
Calibration Manual
34
Lower Display 3-Wire RTD Measurement
1. Press on the UUT (on the 726) until Ω appears on the lower display. If
necessary, use to select the measure mode, and use (on the 726) to get
to the 3W mode. (MEASURE should also appear on the lower display.)
2. Set the 5520A to 2-wire output with 2-wire compensation off; then make the
connections shown in Figure 9.
3. Set the 5520A to source 350 Ω and verify that the UUT resistance readings are
within the 349.80 to 350.2 Ω.
4. Press STBY on the 5520A.
aal10f.eps
Figure 9. Lower Display 3-Wire Resistance Test Connections