Fluke 725 [44/80] Lower display thermocouple source test

Fluke 726 [44/80] Lower display thermocouple source test
724/725/726
Calibration Manual
36
Lower Display Thermocouple Source Test
1. Set the 5520A to measure Type-J thermocouple voltages.
2. Press on the UUT until SOURCE appears on the lower display. If necessary,
press on the UUT until J appears on the lower display and press (use the
configuration menu on the 726) so the temperature is displayed in °C.
3. Use the arrow keys to set the UUT outputs to the temperatures in Table 14 and verify
that the 5520A temperature readings are within the limits shown.
4. Press STBY on the 5520A. Remove the TC lead from the 5520A and the UUT.
Table 14. Lower Display Thermocouple Source Readings
UUT Settings
724/725
5520A Readings
726
Readings
(referenced to 0 °C)
0.0 °C -0.7 °C to +0.7 °C -0.4 °C to +0.4 °C