Fluke 725 [35/80] Lower display mv tc measurement tests

Fluke 726 [35/80] Lower display mv tc measurement tests
Calibrators
Performance Tests
27
Lower Display mV/TC Measurement Tests
1. Press RESET on the 5520A.
2. Press on the UUT until MEASURE and mV appear on the lower display.
3. Make the connections shown in Figure 3.
4. Set up the 5520A to output each of the voltages in Table 8 and verify that the UUT
readings are within the limits shown.
5. Press STBY on the 5520A.
aal04f.eps
Figure 3. Lower Display mV and Voltage Test Connections
Table 8. Lower Display mV Readings
5520A Outputs 724/725 UUT Readings 726 UUT Readings
0.00 mV -0.02 mV to +0.02 mV -0.0100 mV to +0.010 mV
45.00 mV 44.97 mV to 45.03 mV 44.986 mV to 45.014 mV
89.00 mV 88.96 mV to 89.04 mV 88.981 mV to 89.019 mV