724/725/726
Calibration Manual
60
Calibrating mA Input (Fluke 725 and 726 only)
Connect the UUT as shown in Figure 26.
aal06f.eps
Figure 26. mA Input Calibration Connections
After making the proper connections, type “3” then press the Enter key on the PC
keyboard. Input directions are displayed at the bottom of the calibration menu:
Enter Selection:
Input 0 mA, Press Enter When Stable
Input 24 mA, Press Enter When Stable
Use the 5520A to input 0 mA and then 24 mA.