Castellini SKEMA 6 [20/95] Skema 6 skema

Castellini SKEMA 6 [20/95] Skema 6 skema
20
SKEMA 6 / SKEMA 8
- OPERATING INSTRUCTION
20
EN
5.1.1.2.1. AUTOSTERIL disinfection cycle setting
Thissettingissharedbyallusers.
PerformthefollowingstepsfromtheGENERALSETTINGSmenu:
•Touchtheiconbutton
toaccesstheAUTOSTERILDISINFECTION
CYCLESETTINGsubmenu.
NOTE: this submenu can be entered also by pushing at least for
2 seconds the AS key on the assistant’s module.
NOTE: This submenu cannot be entered if the disinfectant liquid
tank is low (see paragraph 7.4.), an instrument is removed or the M.W.B.
system is in an error state. An acoustic signal (BEEP) will signal the
impossibility to enter the submenu.
•Setthedisinfectantdwelltimebytouchingtheiconbuttons
or .
NOTE: the time may range from at least 5 minutes to at most 30
minutes, with 30 second intervals.
WARNING!
Recommended permanence time with PEROXY Ag+: 10 minutes.
Recommended time 3% hydrogen peroxide (10 volumes) should be
left in: 10 minutes.
It is strongly unadvisable to let the oxygenated water dwell in the
ducts for a contact time of more than 30 minutes.
•Withdrawtheinstrumentstobetreated(thecorrespondingiconwillappear
onthedisplay):
S1:syringeoninstrumentboard.
A:instrumentinpositionA
B:instrumentinpositionB
C:instrumentinpositionC
D:instrumentinpositionD
S2:syringeonassistant’sboard.
F:instrumentonassistant'sboard.
CA:suctiontubes.
BC:watertocupduct.
NOTE: pressing the you can select/deselect disinfection
of the cup water duct.
NOTE: if the ushing system of suction cannulas is present, it is
possible to select their ushing by simply putting them into the specially
provided connect couplers (see paragraph 7.5.).
•Tostartthedisinfectioncycle,touchtheiconbutton
(seeParagraph
7.4).
NOTE: the disinfection cycle can be started also by briey pushing
the AS key on the assistant’s module.

Содержание