Fluke 43B [41/54] Function specifications
![Fluke 43B [41/54] Function specifications](/views2/1005303/page41/bg29.png)
Specifications
Function Specifications
3
37
Function Specifications
For all specifications, probe specifications must be added.
Electrical functions
Specifications are valid for signals with a fundamental between 40 and 70 Hz.
Minimum input voltage .................................................................. 4 V peak-peak
Minimum input current.................................................. 10 A peak-peak (1 mV/A)
Input bandwidth.................................. DC to 15 kHz (unless specified otherwise)
Volts / Amps / Hertz
Readings.................................................... V rms (AC+DC), A rms (AC+DC), Hz
Voltage ranges (auto) ........................................... 5.000 V to 500.0 V, 1250 V
±(1 % + 10 counts)
Current ranges (auto) ........................................50.00 A to 500.0 kA, 1250 kA
±(1 % + 10 counts)
Frequency range .............................................................. 10.0 Hz to 15.0 kHz
40.0 to 70.0 Hz ............................................................ ±(0.5 % + 2 counts)
CF Crest Factor range ........................................... 1.0 to 10.0 ±(5 % + 1 counts)
Power
(1 and 3 Phase, 3 Conductor balanced loads)
Readings.................................................................Watt, VA, VAR, PF, DPF, Hz
Watt, VA, VAR ranges (auto)............... 250 W to 250 MW, 625 MW, 1.56 GW
when selected: total (%r): ±(2 % + 6 counts)
when selected: fundamental (%f): ±(4 % + 4 counts)
DPF .............................................................................................. 0.00 to 1.00
0.00 to 0.25............................................................................. not specified
0.25 to 0.90....................................................................................... ± 0.04
0.90 to 1.00....................................................................................... ± 0.03
PF .....................................................................................0.00 to 1.00, ± 0.04
Frequency range .............................................................. 10.0 Hz to 15.0 kHz
40.0 to 70.0 Hz ............................................................ ±(0.5 % + 2 counts)
Harmonics
Number of harmonics......................................................DC..21, DC..33, DC..51
Readings / Cursor readings
V rms / I rms............................................................. fund. ±(3 % + 2 counts)
31
st
±(5 % + 3 counts), 51
st
±(15 % + 5 counts)
Watt.......................................................................... fund. ±(5 % + 10 counts)
31
st
±(10 % + 10 counts), 51
st
±(30 % + 5 counts)
Frequency of fundamental................................................................ ± 0.25 Hz
Содержание
- Fluke 43b 1
- Power quality analyzer 1
- Users manual 1
- Placing orders and getting assistance 2
- Chapter title page 3
- Table of contents 3
- Fluke 43b 4
- The following items are included in your fluke 43b kit 5
- Unpacking 5
- Fluke 43b 6
- Safety information read first 6
- Warning 6
- Safety information read first 7
- Warning 7
- Current probe 8
- Fluke 43b 8
- Warning 8
- Chapter 1 9
- Introducing the fluke 43b 9
- Powering the fluke 43b 9
- Fluke 43b 10
- Inputs 11
- Introducing the fluke 43b input 11
- Fluke 43b 12
- Main menu 12
- Introducing the fluke 43b main men 13
- Fluke 43b 14
- Harmonics 14
- Volts amps hertz 14
- Inrush current 15
- Introducing the fluke 43b main men 15
- Sags swells 15
- Transients 15
- Fluke 43b 16
- Time resolution in sags swells and record modes 16
- Introducing the fluke 43b main men 17
- Fluke 43b 18
- Instrument setup 18
- Setting the date 18
- Accept the new time 19
- Adjust the hours 19
- Introducing the fluke 43b instrument setu 19
- Repeat steps 4 and 5 for minutes and seconds 19
- Setting the time 19
- Adjusting the contrast 20
- Fluke 43b 20
- Selecting probes 21
- Fluke 43b 22
- Move to the probe settings for input 2 22
- Users manual 22
- Introducing the fluke 43b instrument setu 23
- Setting up the harmonic and power function 23
- Fluke 43b 24
- Introducing the fluke 43b instrument setu 25
- Selecting a language 25
- Fluke 43b 26
- Using flukeview software 26
- Introducing the fluke 43b using a printe 27
- Using a printer 27
- Fluke 43b 28
- Selecting a printer type 28
- Resetting the fluke 43b 29
- Chapter 2 31
- Cleaning and storage 31
- Cleaning the current probe 31
- Cleaning the fluke 43b 31
- Maintenance 31
- Storing the fluke 43b 31
- Batteries 32
- Charging the batteries 32
- Fluke 43b 32
- Extending battery operation time 33
- Fluke 43b 34
- Replacing the ni mh battery pack 34
- Warning 34
- Fluke 43b 36
- Ordering codes 36
- Replaceable parts 36
- Maintenance ordering codes 37
- Manuals 37
- Fluke 43b 38
- Flukeview does not recognize the fluke 43b 38
- The batteries operate less than four hours 38
- The fluke 43b does not start up 38
- The printer does not print 38
- The screen remains black 38
- Troubleshooting 38
- Chapter 3 39
- Introduction 39
- Specifications 39
- Fluke 43b 40
- Safety specifications 40
- Electrical functions 41
- Function specifications 41
- Specifications function specifications 41
- Fluke 43b 42
- Specifications function specifications 43
- Fluke 43b 44
- Record 45
- Specifications function specifications 45
- Fluke 43b 46
- Miscellaneous 46
- Environmental conditions 47
- Specifications environmental conditions 47
- Disturbance with test leads tl24 in scope mode v ac dc true rms 48
- Disturbance with test leads tl24 or current clamp i400s volts amps hertz resistance capacitance power harmonics 48
- Electromagnetic immunity 48
- Fluke 43b 48
- No visible disturbance 48
- Ranges not specified in tables 2 and 3 may have a disturbance of more than 10 of full scale 48
- Table 1 48
- Table 2 48
- Table 3 48
- The fluke 43b including standard accessories conforms with the eec directive 89 336 for emc immunity as defined by iec1000 4 3 with the addition of the following tables 48
- Users manual 48
- Current probe i400s 49
- Electrical specifications 49
- Electromagnetic compatibility emc 49
- Safety specifications 49
- Fluke 43b 50
- General specifications 50
- Declaration of conformity 51
- Fluke 43b 52
- Limited warranty limitation of liability 52
- Users manual 52
Похожие устройства
- Indesit IWUD 4105 Инструкция по эксплуатации
- Geo-Fennel FL 60 Maxi-Liner 580000 Инструкция по эксплуатации
- Fluke 51 II Инструкция по эксплуатации
- Indesit IWD 5105 Инструкция по эксплуатации
- LG MS-1744WB Инструкция по эксплуатации
- Blue Weld Prestige Plasma 54 Kompressor Инструкция по эксплуатации
- Supra SWM-774 Инструкция по эксплуатации
- Fluke 52 II Инструкция по эксплуатации
- LG MS-1744W Инструкция по эксплуатации
- Indesit WITL 867 Инструкция по эксплуатации
- Geo-Fennel FL 45 HP 520510 Инструкция по эксплуатации
- Supra SWM-770 Инструкция по эксплуатации
- Belkin F9S600 Инструкция по эксплуатации
- Fluke 53 II Инструкция по эксплуатации
- Indesit WISE 10 CSI Инструкция по эксплуатации
- LG MS-1744U Инструкция по эксплуатации
- Bosch PHC 9690 ProSalon Инструкция по эксплуатации
- Geo-Fennel Square Linner II 551000 Инструкция по эксплуатации
- Supra SWM-755 Инструкция по эксплуатации
- LG MS-1744HL Инструкция по эксплуатации