Fluke 744 [114/138] Specifications

Fluke 744 [114/138] Specifications
744
Users Manual
104
80T-150U Temperature Probe
80PK series thermocouples
80i-410 Clamp-on DC/AC Current Probe
80i-1010 Clamp-on DC/AC Current Probe
80i-500s Clamp-on AC Current Probe (requires
the Y9108 adapter)
80i-1000s Clamp-on AC Current Probe (requires
the Y9108 adapter)
80i-kW Current and Power Probe
Specifications
All specifications apply from +18 °C to +28 °C unless
stated otherwise.
All specifications assume a 5 minute warmup period.
Measurement specifications are valid only when
Damping is turned on. When damping is turned off,
or when the
g annunciator is displayed, floor
specifications are multiplied by 3. Floor specifications
are the second part of the specifications, usually
expressed as "% of full scale." The measure
pressure, temperature, and frequency functions are
specified only with damping on.
The standard specification intervals for the 744 are 1
and 2 years. Typical 90-day source and
measurement accuracy can be estimated by dividing
the 1 year "% of Reading" or "% of Output"
specifications by 2. Floor specifications, expressed
as "% of f.s.", remain constant.
To achieve the best noise rejection, use battery
power and tie all three common jacks together.