Fluke 744 [33/138] Preserving battery life

Fluke 744 [33/138] Preserving battery life
Documenting Process Calibrator
Setting Up the Calibrator
23
Preserving Battery Life
An optional Auto Battery Save feature turns the
calibrator off after a selected idle time. The default
setting for
Auto Battery Save is Off. When Auto
Battery Save
is On, and you are using the optional Ni-
Cd battery pack (Model BP7217), the E symbol
shows in the upper right corner of the display. The
setting is preserved after you turn off the power. Auto
Battery Save works the same when using the battery
eliminator. Turn on the
Auto Battery Save feature as
follows:
1. Press s.
2. Press d to highlight
Off following Auto Battery
Save
.
3. Press e or the
Choices softkey.
4. Press to u highlight
On, then press e.
5. To accept the timeout period shown on the
display, you can finish here. Press
Done to exit
Setup mode and do not go on to step 6.
6. To change the timeout period, press d to
highlight the timeout period following
Battery
Save Timeout
.
7. Press eor the
Choices softkey.
8. Enter your choice of timeout period in minutes
(accepted range: 1 to 120 minutes).
9. Press the
Done softkey.
10. Press the
Done softkey or sto exit Setup
mode.
Using the Optional Battery Eliminator
Caution
To avoid damage to the calibrator, use
only Fluke Model BE9005 Series Battery
Eliminator, available from your Fluke
representative.