Fluke 744 [71/138] Using auto step

Fluke 744 [71/138] Using auto step
Documenting Process Calibrator
Using Source Mode
61
Using Auto Step
To have the calibrator make a sequence of steps
automatically, either once through the sequence or
repetitively, proceed as follows:
1. Refer to the appropriate “Using Source Mode”
subheading earlier in this manual (e.g.,
“Sourcing Electrical Parameters”) and connect
the calibrator to the circuit to be tested.
2.
If necessary, press M for SOURCE mode.
3.
Set the calibrator for the desired source value.
4.
If you want to step the source value in % of
scale, set the % of scale value as described
previously under “Sourcing in Percent of Scale.”
5.
Press the
Step softkey.
6.
Press the
Auto Step softkey.
7.
Select values for the following parameters as
you are prompted by the display:
Starting point (in units or % of scale)
Ending point
Number of steps
Time per step
Single shot or continuous repetition
Ramp pattern if continuous
Start delay
8.
To start automatic stepping, press the
Start Step
softkey. The softkey label changes to Stop Step.
9.
To stop automatic stepping, press the
Stop Step
softkey.
10.
Press the
Done softkey to resume normal
operation.