Fluke 744 [44/138] Using thermocouples

Fluke 744 [44/138] Using thermocouples
744
Users Manual
34
Measuring Temperature
Using Thermocouples
The calibrator supports eleven standard
thermocouples, each identified with an alpha
character: E, N, J, K, T, B, R, S, C, L, or U. Table 5
summarizes the ranges and characteristics of the
supported thermocouples.
To measure temperature using a thermocouple,
proceed as follows:
1. Attach the thermocouple leads to the appropriate
TC miniplug, then to the TC input/output as
shown in Figure 13.
One pin is wider than the
other. Do not try to force a miniplug in the wrong
polarization
.
Note
If the calibrator and the thermocouple plug
are at different temperatures, wait one
minute or more for the connector
temperature to stabilize after you plug the
miniplug into the TC input/output.
2. If necessary, press M for MEASURE mode.
3. Press t. The display prompts you to select
the thermocouple type.
4. Select the desired thermocouple type using the
u or d followed by e.
5. If necessary, you can change between
°C or °F
Temperature Units
as follows:
a. Press s.
b. Press the
Next Page softkey twice.
c. Use the u and d keys to move the cursor
to the desired parameter. Then press either
eor the
Choices softkey to choose a
setting for that parameter.
d. Press u or d to move the cursor to the
desired setting.
e. Press eto go back to the sdisplay.
f. Press the Done softkey or sto exit Setup
mode.
6. If necessary, you can change between
ITS-90 or
IPTS-68 Temperature Scale in Setup mode. The
procedure is the same as steps a-f above.

Содержание