Cleaning
The atomiser and the accessories such
as mouthpiece, mask, nosepiece, nozzle
attachment, etc. must be cleaned with hot
water after each use. Carefully dry all
components using a soft cloth. Once they
are completely dry, reassemble the parts
and place in a dry, sealed container or
proceed with sterilisation.
When cleaning, make sure that all resi-
dues are removed. Never use any sub-
stances that may be poisonous when in
contact with the skin or mucous mem-
branes, or when swallowed or inhaled.
Important
x Before you clean the unit, always
switch it off, unplug it and let it cool off.
x Ensure that no water penetrates inside
the unit!
x Do not put the device or accessories in a
dishwasher.
x Do not touch the unit with wet hands
while it is plugged in; do not allow any
water to be sprayed onto the unit.
x The unit must be operated only when it
is completely dry.
x Do not spray any liquids into the venti-
lation slots. Any liquid that penetrates
into the unit can damage the electrical
parts or other components of the com-
pressor and impair the function of the
device.
x Clean the device using a damp cloth, to
which you can apply a little detergent if
necessary.
x Do not use abrasive detergents and ne-
ver immerse the unit in water.
Condensation, care of the hose
Depending on the environmental conditions,
condensation may collect in the hose. To
prevent against bacterial growth and to en-
sure effective therapeutic use, this moisture
must always be removed. To remove mois-
ture from the hose, proceed as follows:
x Remove the hose [15] from the nebu-
liser [11].
x Leave the hose plugged into the com-
pressor [6].
x Run the compressor until all the mois-
ture has been displaced by the flow of
air.
In the event of significant contaminant de-
posits, replace the hose.
Sterilisation
Accessories (with the exception of the ho-
se) can be sterilised using specific prod-
ucts such as the MILTON® method: they
must be rinsed until all traces of disinfec-
tant are removed, dried, and stored in a
dry and dust-free room. Accessories (with
the exception of the hose) can also be ste-
rilised by placing them for at least 40 min-
utes in boiling water (100°C) or at least 20
minutes in an autoclave (120°C). Do not
bend the accessories or place any weights
on top of them. Do not exceed the speci-
fied temperatures, and use a sterile bag if
possible. Ensure that the water level is suf-
ficient to prevent nebuliser parts coming
into direct contact with the bottom of the
pan.