Fluke 726 [26/80] Hart resistor on off

Fluke 726 [26/80] Hart resistor on off
726
Users Manual
16
HART
Resistor ON/OFF
1. Press C until SELECT HART ON or OFF
appears on the display.
2. Use V to toggle ON or OFF.
3. Press S to save the setting.
Note
When HART mode is selected, the 250
resistor is turned on both mA channels.
Getting Started
This section details some basic operations of the
Calibrator.
Voltage to Voltage Test
To perform a voltage-to-voltage test:
1. Connect the Calibrator’s voltage output to its voltage
input as shown in Figure 6.
2. Press
O to turn on the Calibrator. Press l to
select dc voltage (upper display).
3. If necessary, press
M for SOURCE mode (lower
display). The Calibrator is still measuring dc voltage,
the active measurements are visible in the upper
display.
4. Press
V to select dc voltage sourcing.
5. Press Y and Z to select a digit to change. Press X
to select 1 V for the output value. Press and hold
J
to enter 1 V as the 0 % value.
6. Press X to increase the output to 5 V. Press and
hold
G to enter 5 V as the 100 % value.
7. Press
H and I to step between 0 and 100 %
in 25 % step increments.