Fluke 726 [70/80] Ohms measurement

Fluke 726 [70/80] Ohms measurement
726
Users Manual
60
Ohms Measurement
Ohms Range Minimum Maximum
Accuracy
(% of Reading + Floor)
Ohms Read (low) 0.00 400.00
0.015 % + 0.05
Ohms Read (high) 401.0 4000.0
0.015 % + 0.5
Ohms Source
Ohms Range Minimum Maximum
Excitation Current from
Measurement Device
Accuracy
(% of Reading + Floor)
5.0 400.0 0.1 to 0.5 mA
0.015 % + 0.1
Ohms Source (low)
5.0 400.0 0.5 to 3 mA
0.015 % + 0.05
400 1500 0.05 to 0.8 mA
0.015 % + 0.5
Ohms Source (high)
1500 4000 0.05 to 0.4 mA
0.015 % + 0.5
Unit is compatible with smart transmitters and PLCs.
Frequency response is 5 mS
Frequency Measurement
Range Minimum Maximum
Accuracy
(% of Reading + Floor)
CPM Read 2.0 1000.0 0.05 % + 0.1 CPM
Hz Read 1.0 1000.0 0.05 % + 0.1 Hz
KHz Read 1.00 15.00 0.05 % + 0.01 KHz