Type
Minimum Maximum
CJC ON
Accuracy
CJC OFF
Accuracy
-20 0.0 2.0 1.8
S
0.0 1767 1.4 1.2
600 800 1.4 1.2
800 1000 1.5 1.3
B
1000 1820 1.7 1.5
0.0 1000 0.8 0.6
C
1000 2316 2.5 2.3
-200 0.0 0.45 0.25
L
0.0 900 0.4 0.2
-200 0.0 0.7 0.5
U
0.0 600 0.45 0.25
-200 0.0 1.0 0.8
N
0.0 1300 0.6 0.4
XK -200 800 0.4 0.2
0.0 800 1.1 0.9
BP
800 2500 2.3 2.1
Range Accuracy
Thermocouple in mV read
-10 °C to 75 °C
0.015 % + 10 µV
(% of Reading + Floor)
Thermocouple in mV source
-10 °C to 75 °C
0.015 % + 10 µV
(% of Reading + Floor)
Maximum current output in voltage ranges is 1 mA with an output impedance of 1 e