Fluke 726 [40/80] Using source mode

Fluke 726 [40/80] Using source mode
726
Users Manual
30
Using Source Mode
In SOURCE mode, the Calibrator:
generates calibrated signals for testing and
calibrating process instruments
supplies voltages, currents, frequencies, and
resistances
simulates the electrical output of RTD and
thermocouple temperature sensors
measures gas pressure from an external source,
creating a calibrated pressure source.
Sourcing 4 to 20 mA
To select the current sourcing mode, proceed as follows:
1. Connect the test leads in the mA terminals (left
column).
2. If necessary, press
M for SOURCE mode.
3. Press
V for current and enter the desired current
by pressing X,W,
Y, and Z.
Simulating a 4- to 20-mA Transmitter
Simulate is a special mode of operation in which the
Calibrator is connected into a loop in place of a
transmitter and supplies a known, settable test current.
Proceed as follows:
1. Connect the 24 V loop power source as shown in
Figure 14.
2. If necessary, press
M for SOURCE mode.
3. Press
V until both mA and SIM display.
4. Enter the desired current by pressing X,W,
Y, and
Z.