Users Manual
78
16. Set
Trip Function to Trip Cont by cycling through
the choices with the e key.
17. Press the
Done softkey.
18. Press the
Manual Test softkey.
19. Close the pressure line vent and slowly bring the
pressure up to the trip point.
20. When the switch sets, slowly bring the pressure
back down until the switch resets. You can
repeat this cycle as many times as you want.
21. Press the
Done softkey and view the results.
22. Press the
Done softkey and if desired, enter Tag,
S/N,
and/or ID.
23. Press the
Done softkey.
24. Now press the
Adjust softkey if you want to
adjust the limit switch and test it again.
25. Use the softkeys to control the calibrator, and
adjust the limit switch as necessary.
26. Press the
Done softkey.
27. Press the
As Left softkey to run the test again
with the same parameters. Results from the As
Found and As Left tests are saved in the
calibrator memory for later viewing or uploading.
The procedure for limit switches that respond to
other parameters work similarly. When you do a 2 Pt.
Limit Switch Test, you simply follow the prompts on
the display for testing the first switch, changing test
leads, and testing the second limit switch.