Documenting Process Calibrator
Calibrating a Process Instrument
77
The procedure for testing a pressure limit switch
follows. The switch in this example sets at a high limit
of 10 psi. The set state is a closed switch contact.
For pressure switches, you use the
Manual Test
choice. For testing switches that do not require
sourcing pressure, you can use the
Auto Test choice.
1. Connect the test leads between the pressure
switch contact output and the mA RTD
(middle) jacks on the calibrator.
2. Connect the pressure module to the calibrator,
and connect a pressure line to the limit switch.
Leave the pressure line vented to atmosphere.
3. If necessary, press M for MEASURE mode.
4. Press q q for the continuity measure
function.
5. Press M for SOURCE mode.
6. Press p for the pressure source function.
7. Press c to zero the pressure module.
8. Press M.
9. Press the
As Found softkey.
10. Highlight
1 Pt. Switch Test from the menu and
press e.
11. Press e to modify the parameters for
Setpoint 1.
12. Make the following selections:
Setpoint 1 = 10.000 psi
Setpoint Type = High
Set State = Short
13. Press the Done softkey.
14. Set the
Tolerance to 0.5 psi.
15. The next parameters,
Deadband Min and
Deadband Max, are optional. Do not set them in
this example.