And ВМ-22 [41/88] Id id number output

And ВМ-300 [41/88] Id id number output
BM series 39
Class Item and Parameter Description
0
Every measurement
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minutes
6
2 minutes
7
5 minutes
int
Interval time
8
10 minutes
Interval time in the interval memory
mode when using
prt
3,
,
data
1
off
No output
d-no
Data number output
on
Output
Refer to "
12. Data Memory
"
0
No output
1
Time only
2
Date only
5-td
Time/Date output
3
Time and date
Selects whether or not the time or date
is added to the weighting data. Refer to
"
10.7. Clock and Calendar
Function
" for details.
off
No output
5-id
ID number output
on
Output
Selects whether or not the ID number
is output.
off
No pause
pU5e
Data output pause
on
Pause (1.6 seconds)
Selects the data output interval.
off
Not used
at-f
Auto feed
on
Used
Selects whether or not automatic feed
is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
Refer to "
11. ID Number And GLP
Report
" for details.
off
Not used
dout
Data output
ar-d
Zero after output
on
Used
Adjusts zero automatically after data is
output
0
600 bps
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
bp5
Baud rate
5
19200 bps
0
7 bits, even
1
7 bits, odd
btpr
Data bit, parity bit
2
8 bits, none
0
CR LF
Crlf
Terminator
1
CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
5if
Serial
interface
type
Data format
5
CSV format
Refer to "
10.5. Description of Item
"Data Format"
".
:
Factory settings.
Caution The balance may not transmit the data completely at the specified refresh rate, depending
on the baud rate or data added to the weighing data such as time, date and ID number.

Содержание