48 BM series
11. ID Number And GLP Report
The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The GLP output format is selected at "GLP output (info)" of the function table and can be output
to a personal computer or printer using the RS-232C serial interface.
The GLP output format includes the balance manufacturer, model, serial number, ID number,
date, time and space for signature.
The balance can output the following reports for GLP using the RS-232C serial interface.
"Calibration report" of the "Automatic self calibration" and "One-touch calibration" using the
internal mass.
"Calibration report" of the calibration using an external weight.
"Calibration test report" of the calibration test using an external weight.
"Title block" and "End block" for the weighing data.
Calibration and calibration test data can be stored in memory and several reports can be output at
the same time. Refer to "12. Data Memory" for details.
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
Refer to "10.7. Clock and Calendar Function" for checking and storing the time and date.
11.1. Setting The ID Number
1 Press and hold the RANGE key until ba5fnc of the function table is displayed, then release
the key.
2 Press the RANGE key several times to display id .
3 Press the PRINT key. Set the ID number using the following keys.
RE-ZERO key
.....
The key to set the character of the digit selected.
Refer to the display character set shown below.
RANGE key
........
The key to select the digit to change the character.
PRINT key
..........
The key to store the new ID number and display ba5fnc .
CAL key
.............
The key to cancel the new ID number and display ba5fnc .
4 When displaying
ba5fnc , press the CAL key to return to the weighing mode.