And ВМ-22 [42/88] Description of the class environment display

And ВМ-252 [42/88] Description of the class environment display
40 BM series
off
No limit
t-Up
Timeout
on
1 second
Selects the wait time to receive a
command.
off
No output
5if
Serial
interface
erCd
AK, Error code
on
Output
AK: ASCII code 06h
0
Water temperature
d5
fnc
Density function
ldin
Liquid density input
1
Liquid density
Available only when density mode is
selected
Unit
Unit
Refer to "5. Weighing Units".
C5 in
Internal mass correction
Displayed only when the internal mass value correction switch is
set to 1. Refer to "8. Calibration
(To Adjust The Weighing Value)".
id
ID number setting
Refer to "11. ID Number And GLP Report".
0
3 seconds
1
10 seconds
ionfnc
Static eliminator
ion
Neutralization time
2
Manual operation (max. 10 min.)
Refer to "
6.5. Built-in DC Static
Eliminator"
".
off
No used
door
Door state function
d-dt
Door sensor
on
Used
:
Factory settings.
10.3. Description Of The Class "Environment, Display"
Condition ( Cond )
Cond 0 This parameter is for sensitive response to the fluctuation of a mass value. Used for
powder target mass, weighing a very light sample or when quick response weighing is
required. After setting, the balance displays FAST.
Cond 2 This parameter is for stable weighing with slow response. Used to prevent a mass
value from drifting due to vibration or drafts. After setting, the balance displays SLOW.
Notes In automatic response adjustment, this parameter is selected automatically.
Stability band width ( 5t-b )
This item controls the width to regard a mass value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs or
stores the data. The parameter influences the "Auto print mode".
5t-b 0 This parameter is used for sensitive response of the stabilization indicator. Used for
exact weighing.
5t-b 2 This parameter ignores slight fluctuations of a mass value. Used to prevent a mass
value from drifting due to vibration or drafts.
Zero tracking ( trc )
This function tracks zero point drift caused by changes in the environment and stabilizes the zero
point. When the weighing data is only a few digits, turn the function off for accurate weighing.
trc 0 The tracking function is not used. Used for weighing a very light sample.
trc 1 The normal tracking function is used.
trc 2 The strong tracking function is used.
trc 3 The very strong tracking function is used. Used for stable zero display.

Содержание