15.3. Commands
15.3.1. Command List
Note A command has a terminator added, that is specified using "5if, Crlf" of the function
table, and is sent to the balance.
Commands to query weighing data
C Cancels the S or SIR command.
Q Requests the weighing data immediately.
S Requests the weighing data when stabilized.
SI Requests the weighing data immediately.
SIR Requests the weighing data continuously.
Commands to control the balance
CAL
Same as the 1CAL1 key. Calibration using the internal mass.
OFF
Turns the display off.
ON
Turns the display on.
P
Same as the ON:OFF key. Turing the display on or off.
PRT
Same as the
PRINT key.
R
Same as the RE-ZERO key. Zero display.
#1
RNG
Same as the RANGE key. Changing the minimum display.
TR
Tares the balance. Canceling the container’s weight..
#1
TST
Perform calibration test.. Calibration test using the internal mass.
U
Same as the MODE key. Changing the unit.
Commands to request data
?ID
Requests the identification number.
?PT
Outputs the tare value.
#1
?SN
Request the serial number of the balance.
?TN
Request the model name of the balance.
Command to set data
PT:∗∗∗.∗∗∗∗ g
Sets the tare value.
#1
#1
Command R assumes the point as zero and sets the display to zero.
Command TR cancels the tare weight when the displayed value is greater than zero.
The canceled tare value can be read using the command ?PT.
Command PT:∗∗∗.∗∗∗∗ g sets the tare value digitally.
An example of this command sets a negative target value and fills with the sample until the
display becomes zero. For details, refer to page 71.
Step 1
Place a container.
Step 2
Set the display to zero using the R command.
Step 3
Set a target value using the PT: command.