2-2. Inspection of Endoscope
CAUTION:
If the endoscope is intended to be used clinically after testing of individual endoscope
functions without further reprocessing, the following precaution should be exercised.
Use sterile or bacterial free water during individual endoscope function tests to
avoid recontamination of the previously reprocessed instrument by waterborne
microorganisms. Tap water, especially that which may be left idle and uncovered for
a prolonged period of time, should not be used during any inspection/testing of the
endoscope.
Before reprocessing and/or immersion in any uids, PENTAX endoscopes should
be tested for the loss of integrity in their watertight construction by using PENTAX
brand leak testers. For specic details on PENTAX recommended leak detection
procedures, please refer to the instructions supplied with PENTAX leak testers.
CAUTION:
Various types of endoscope leak testers exist including manual, electro-mechanical
and “automated” versions, some of which are stand alone units and others which may
be integrated into Automated Endoscope Reprocessors (AERs)/Washers-Disinfectors
(WDs). It must be recognized that PENTAX does not evaluate non-PENTAX leak
testersystemstosatisfytheirspecicproductclaims,fortheireffectivenessto
accurately detect leaks, and/or for their compatibility with PENTAX endoscopes.
Insufcientpressuresmayreducethelikelihoodforaccurateleakdetection,especially
iftheendoscope'sdistalbendingsectionisnotexedduringtesting.Excessive
pressures may adversely affect the endoscope, especially if pressurization occurs
during automated reprocessing at elevated temperatures. PENTAX accepts no
responsibility for use of non-PENTAX leak testers.
Users should check with the leak tester manufacturer and confirm their specific
product claims, including compatibility with PENTAX endoscopes at various
temperaturesandtheirabilitytodetectleakswith/withoutuidimmersionandwith/
withoutexingoftheendoscope'sdistalbendingsection.
1) Inspection of the Insertion Tube
a) Check the entire surface of the insertion tube for abnormal conditions such as
dents, crush marks, wrinkles, bumps, buckles, excessive bending, protrusions,
peeling of outer sheath, cuts/holes or other irregularities. Any crush or
indentation of the flexible shaft of the endoscopes can cause damage to the
internal mechanisms of the endoscopes.
b) Similarly, check the condition of the umbilical cable for outward signs of damage
such as buckling, crush marks, etc.