– 55 –
4-2-5-1. ETO
WARNING:
●Prior to sterilization, clean and dry the endoscope thoroughly. Failure to thoroughly
dry all surface areas could result in incomplete or ineffective sterilization. Moisture
could prevent contact of the ETO gas with the actual contaminated surfaces.
●Use appropriate heat process indicators and/or biological monitors as
recommended by the manufacturer of the sterilizer.
●After the sterilization, ensure that the package is intact. If there are any signs of
abnormalities such as stains, tears or any other indications that the packaging
has previously been damaged or opened, repeat the sterilization with a new
packaging.
Ethylene Oxide (ETO) Gas Sterilization can be performed on these Endoscopes,
provided the following special instructions are followed to ensure the proper
performance of the instrument. Adhere to the sterilizer manufacturers instructions and
always use chemical and/or biological indicator.
4-2-5-1-1. Preparation
CAUTION:
Prior to the ETO sterilization, ensure the ETO (OF-C5) cap is attached to the
endoscope. Failure to do so could result in burst of the bending rubber.
1) Ensure the cleaning cap (OF-B155) and the rubber inlet seal (OF-B190) are
removed from the endoscope.
(1)
(1)
(1) Detach the cleaning
cap (OF-B155)
(2) Detach the rubber inlet
seal (OF-B190)
Figure 4.18
2) Remove the soaking cap (OE-C9) from the endoscope.
3) Attach the ETO cap (OF-C5) to the endoscope.
(1)
(2)
(1) Attach the ETO Cap
(2) Detach the PVE
Soaking Cap
Figure 4.19