Pentax ECY-1575K [61/80] 2 5 3 other methods of sterilization

Pentax ECY-1575K [61/80] 2 5 3 other methods of sterilization
– 61 –
4-2-5-3. Other Methods of Sterilization
Other types of cleaning, disinfecting and/or sterilization systems/processes are available
for the reprocessing of medical devices. However, due to the heat sensitive nature and/or
the specic biocompatible materials used in the construction of exible endoscopes, some
of these marketed systems/processes/solutions could have detrimental effects on flexible
endoscopes. To avoid the potential for instrument damage and/or endoscope failure, conrm
the compatibility of such reprocessing systems/solutions with your local PENTAX dealer
prior to use with any PENTAX products. Prior to using other methods, conrm the specic
claim(s) of any sterilization methods/processes with the sterilizer manufacturer and ensure
manufacturer that they have performed microbiological validation studies that support
their claims of achieving sterilization of device specic exible endoscopes and endoscope
components.
WARNING:
Inadditiontoefcacyclaimsalwayscheckwiththemanufacturerofthesterilizerto
conrmthattheyhavetestdatademonstratingthatnoharmfullevelsofanyresidues
(active/inert ingredients, their byproducts or derivatives of the processed devices)
remain on any instrument or lumened surfaces that may pose a risk to patients and
users.