– 43 –
4-2-3. Cleaning
WARNING:
●Immediately (at least within one hour) after the completion of a procedure, the
endoscope and its components should be thoroughly and carefully cleaned with
detergent solution. If the endoscopes and its components are left uncleaned
for some time after use, dried blood, mucous or other patient debris may cause
damage or interfere with the ability of the user to proper reprocessing.
●It is extremely important to use detergent solution immediately after each
procedure to dissolve and remove organic contaminants and proteinaceous.
Failure to do so could result in incomplete or ineffective disinfection or sterilization.
●For cleaning, use only legally marketed detergent which has been tested
according to the instructions of the manufacturer and found to be compatible by
PENTAX.Forspecicbrandsofcompatibledetergent,pleasecontactyourlocal
PENTAX service facility or sales representative.
●Please replace the detergent solution for each patient use from the standpoints
of infection control and functionality.
CAUTION:
●The PVE Soaking cap (OE-C9) should be securely ON the electrical contacts.
Failure to do so could result in water invasion and damage to the endoscope.
When the endoscope was erroneously cleaned without the soaking cap attached,
do not use the endoscope and contact your local PENTAX service facility or sales
representative.
●The 'Red"ETO gas sterilization venting cap (OF-C5) must be taken OFF.
Failure to do so could result in water invasion and damage to the endoscope.
When the endoscope was erroneously cleaned with the ETO cap attached, do
not use the endoscope and contact your local PENTAX service facility or sales
representative.
●During cleaning, never twist, rotate or bend the insertion portion, and umbilical
cable excessively.
●Never subject the endoscope to ultrasonic cleaning methods.