Pentax ECY-1575K [67/80] 3 2 1 items required for disinfection

– 67 –
4-3-2-1. Items required for disinfection
protective garments such as gloves, gowns, face masks, etc. to minimize the
risk of cross contamination.
disinfecting solution
sterile water or at least clean potable water suitable to drink
70-90% medical grade ethyl or isopropyl alcohol
basin (at least 25 cm in width x 20 cm in depth x 15 cm in height)
sterile gauze
10 mL syringe
4-3-2-2. Disinfecting Procedure
1) Wear the protective garments.
2) Prepare the basin with disinfecting solution per manufacturers instructions.
3) Immerse fully the accessories and the components and keep them immersed in
the disinfecting solution during the following disinfecting procedure.
4) Manipulate the suction valve (OF-B179) and using a 10 mL syringe, inject
disinfecting solution into the lumen.
5) Manipulate the irrigation valve (OF-B163).
6) Still fully immersed, ensure there is no air bubbles on the surfaces of the
accessories and the components. If any, inject disinfecting solution via a 10 mL
syringe to remove the air bubbles.
7) Immerse them in the disinfecting solution for the appropriate amount of time
according to the instructions of the disinfectant manufacturer.
8) Remove them from the disinfecting solution.
4-3-2-3. Rinse
1) Prepare a basin with water and immerse the accessories and the components .
2) Rinse the external surfaces in the basin.
3) Discard the water.
4) Fill the basin with water and repeat the steps 2) and 3) above.
5) Fill the basin with water.
6) Manipulate the suction valve (OF-B179) and using a 10 mL syringe, inject water
3 mL or more into the lumen to remove the residual disinfectant.
7) Manipulate the irrigation valve (OF-B163)
8) Remove the accessories and the components from the disinfecting solution.
9) Using a 10 mL syringe, ush air into the lumen of the suction valve (OF-B179) to
expel the residual water.
10) Using a 10 mL syringe, ush alcohol 1 mL into the lumen of the suction valve
(OF-B179) to facilitate drying.
11) Using a 10 mL syringe, ush air into the lumen of the suction valve (OF-B179).
12) Gently dry all external surfaces of the accessories and the components with sterile
gauze moistened with alcohol.