RUBBER STRAIN RELIEF
ELECTRICAL CONTACTS
PVE CONNECTOR
Can be rotated within a
180° range.
SOAKING CAP (OE-C9)
This cap must be securely attached before immersion. Align
the black arrow on the soaking cap with the green dot at the
base of the silver collar surrounding the electrical contacts
on the PENTAX PVE connector. Press the cap down onto
the metal collar and turn clockwise to secure.
FEED BACK TERMINAL
To connect the OL-Z3 cable
from the PENTAX video
processor, EPK-i series/1000.
See detail information in
section 3-5.
VENTING CONNECTOR
Accepts “RED” ETO GAS Sterilization
Venting Cap.
Also accepts Leakage Tester.
INSERTION TUBE
Provides venting of endoscope interior to equalize internal and
external pressures.This cap must be removed before immersion.
ETO GAS STERILIZATION VENTING CAP RED (OF-C5)
BENDING SECTION
NOTE: See important separate section
regarding the use of this cap!
(p. 40, 47, 54)
– 5 –
NOTE:
Ensure that the soaking cap has been securely attached (by properly rotating it) to
prevent the cap from coming off during reprocessing. Failure to securely attach the
soaking cap can result in endoscope damage.
NOTE:
To avoid damaging the endoscope, do NOT twist, rotate or bend excessively any of
the rubber strain reliefs.
WARNING:
Immediately after use, the metal light guide prong and the electrical contacts/pins of
the endoscope may be HOT. To avoid burns, do not touch these areas immediately
after use. For safer handling after a procedure, grasp the PVE connector housing.