Pentax ECY-1575K Руководство пользователя онлайн [16/80] 419919

Pentax ECY-1575K Руководство пользователя онлайн [16/80] 419919
– 16 –
NOTE:
Flexible endoscopes and other sophisticated medical instruments are constructed
of special materials, unique parts and intricate components with strict dimensional
tolerances.Specializedassemblytechniquesandapplicationofspecicsealants
and/or adhesives are required to ensure the watertight integrity and maintain the
functionality of these devices. It is therefore imperative that endoscopes be routinely
checked to ensure that parts used in their construction are not loose, missing or
compromised that could otherwise negatively affect the functionality of these devices.
Compromised or loose components could result in device failure, endoscope damage
(viauidinvasion)and/orinincompletedecontaminationofusedinstruments.
PENTAX recommends that prior to use endoscopes should be carefully inspected for
their integrity and checked for any “looseness” in the mating or joining of components
including the following parts/areas:
●channel inlet assembly (biopsy inlet port) (1)
●suction/irrigation nipple/connector (2)
●suction/irrigation valve cylinder (3)
●venting connector (4)
●basically, any inlet or outlet port associated with an internal channel, an indirect
patient contact portion of the endoscope
●rubber strain relief along insertion tube and umbilical cable (rotate clockwise only
to tighten)
(1)
(2)
(3)
(4)
Figure 2.8
One method to check for looseness is to lightly grip the exposed part, and while
grasping the component carefully attempt to move it in various directions. Use of a
lint-free gauze while grasping metal parts is recommended as a protection for one’s
ngers.
If any part/component remains loose (after attempting to tighten) and/or if there is any
indication or suspicion of an abnormality or outward signs of damage, do NOT use
the endoscope. Contact your local PENTAX service facility.

Содержание

Скачать