Calibrators
Calibration Adjustment
41
To begin calibration, or to select a particular calibration step, type the cal step (letter or
number), then press Enter on the PC keyboard. It is not necessary to run all of the
calibration steps when in the calibration mode. However, for the 726, certain
dependencies exist in that certain functions must be calibrated before others can be
calibrated. These dependencies are related to calibration only.
You must first calibrate mV Input and mV Output before calibrating thermocouples for the
724, 725, and 726. At the end of each step on the calibration menu, the new calibration
constants for that step are saved, but not actually used until power is recycled.
When calibrating a measurement function, you must enter an input signal. For the 724
and 725, when the signal is connected and stable, press the space bar to start the
adjustment. For the 726, press the Enter key. Usually, the UUT will display a calibration
constant, then prompt for a second input value. Apply the new input value, then press the
space bar or Enter key depending on the Calibrator. After the adjustment is complete,
press the proper key to return to the Calibration Menu.
When calibrating a source function, you must enter the value of a reading. Type in the
numerical value of the calibration desired. This will put the UUT into the desired
calibration mode. When entering the calibration data for any source mode calibration, be
sure to enter the value in the units listed, but don't enter the units (mV, mA, etc.). After
the adjustment is complete, press the proper key to return to Calibration Menu. The
calibration values will vary slightly, unit to unit. The constants used in this procedure may
not appear exactly the same as on your UUT, but they should be approximately the
same.
Later 724 and 725 have the same adjustment instructions as the 726. For 724 or 725
with V3.00 or later firmware, go to the instructions later in the manual for the adjustment
of the 724-726
To find the firmware of a 72x, push and hold 0% when the unit is turned on. Continue to
hold during the 30 V warning, about 5 seconds. The firmware version will then be shown
in the right-hand section of the display. If the unit does not show the 30 V warning, or a
firmware version, it is an early V1.xx firmware.