Calibration Manual
56
Calibration Adjustment Procedures (726 and 724/725 with V3.00 and Higher)
Note
This procedure does not cover the verification of the unit. The Calibrator
should never use up more than 50 % of its specifications immediately after
calibration. When calibrating the 726, you can also exit and enter
calibration mode while performing the verification, subject to the
dependencies listed in the procedure.
Initiating Communication
The subsequent sections detail the calibration procedures for the 726 and 725/724
firmware V3.00 or higher.
To initiate calibration-mode communication, use the terminal to send the CAL_START
command. Enter the required password.
The password for all 726 Calibrators is 627.
The password for all 725 Calibrators is 527.
The password for all 724 Calibrators is 427.
One of the calibration menus (below) will then be shown on the terminal.
726 Calibration Menu
Calibration Menu
1: ISO mA Input
2: ISO Volts Input
3: mA Input
4: mA Output
5: Volts Input
6: Volts Output
7: Low Ohms Source
8: High Ohms Source
9: Ohms Measure
10: TC mV Output
11: TC mV Input
12: TC CJC
13: Exit
Enter Selection: