724/725/726
Calibration Manual
42
Initiating Communication (724 and 725 with V2.xx and Lower Firmware)
Starting with the UUT off, push and hold while turning the UUT on. Continue to hold
until "Cal mode" is displayed. “Enter Password:" appears on the display of all
Calibrators that have firmware version V1.91 or higher. A password has been added to
prevent users from accidentally changing the calibration of the Calibrator. The password
for all 724 Calibrators is 427. The password for all 725 Calibrators is 527.
The calibration menus, as seen on the PC screen, are as follows:
725 Calibration Menu
Calibrate Menu
1 - Cal Volts Input
2 - Cal Volts Output
3 - Cal mA Input
4 - Cal mA Output
5 - Cal mV Input
6 - Cal mV Output
7 - Cal Thermocouples
8 - Cal Ohms Hi Source
9 - Cal Ohms Low Source
A - Cal RTD LOW Range
B - Cal RTD HI Range
C - Cal ISO Volts
D - Cal ISO mA
Enter Selection :
724 Calibration Menu
Calibrate Menu
1 - Cal Volts Input
2 - Cal Volts Output
3 - Cal mV Input
4 - Cal mV Output
5 - Cal Thermocouples
6 - Cal Ohms Hi Source
7 - Cal Ohms Low Source
8 - Cal RTD LOW Range
9 - Cal RTD HI Range
A - Cal ISO Volts
B - Cal ISO mA
Enter Selection :