The test procedure is shown as figure 3-10. The load can real-time display battery voltage, discharge
current, discharge time, and discharge capacity during the test. The maximum battery discharge time is
99hours 99minutes 99 seconds.
1 - Battery Voltage
2 - Constant Voltage Discharge
3 - Discharge Current
4 - Constant Current Discharge
5 - Termination Current
3.7. Short Circuit Operation
The electronic load can simulate a short circuit to test the protection performance of the tested device.
The short circuit operation can be enabled and disabled by setting in the main menu, or via the remote
command (INPut:SHORt ON/OFF). When the short circuit operation is enabled, the relevant short-circuit
value depends on the present mode of the load. See “Main Specifications” section for details.
The other set values will not be changed when the short circuit operation is enabled.
Note:
In short circuit operation, the Von point and current limit level should be taken into consideration, which
may cause the shut down of the input, so that the short circuit operation is interrupted.