AXIOMET AX-EL300W60A [53/72] Saving and recalling

AXIOMET AX-EL300W [53/72] Saving and recalling
4 /// Use “↑”, “↓” keys to select parameters to input termination voltage (Mini Volt). Use Entry keys or the
knob together with ”, keys to input termination voltage to 15V, and press “Enter” key to confirm.
/// MIN VO 15.000V
5 /// Use “↑”, “↓” keys to select parameters to input termination current (Mini Curren). Use Entry keys or
the knob together with “ ”, ” keys to input termination current to 1A, and press “Enter” key to confirm.
/// IOFF P 01.000A
6 /// Use Entry keys or the knob together with ”, keys to input discharge current to 3A, and press
“Enter” key to confirm. /// DISCUR 3.000A
7 /// Press “Cancel” key to exit the battery discharge parameters edit screen. ///
8 /// Press “On/Off” key to turn on the load, and start battery discharge operation. ///
9 /// Press “On/Off” key to turn off the load, and stop battery discharge operation. ///
10 /// Press “Cancel” key to clear the discharged time and discharged capacity of the battery. ///
11 /// Press mode hotkey to exit the battery discharge operation ///
For the above operating procedures, the corresponding SCPI commands should be:
INPUT OFF ; Turns off the load
BATTERY ON ; Activates the battery discharge operation screen
BATT:VOLT:OFF 15 ; Sets termination voltage to 15V
BATT:DISC:CURR 3 ; Sets termination current to 3A
INPUT ON ; Turns on the load, and start battery discharge operation
INPUT OFF ; Turns off the load, and stop battery discharge operation
BATTERY OFF ; Exits the battery discharge operation
5.10. Saving and Recalling
The operating procedures for saving and recalling operation are shown as below:
1) Press “Shift”+“1Store” keys to enter into the saving menu, or press “Shift”+“2Recall” keys to enter into
the recalling menu.
2) Press Entry keys or the knob to select saving or recalling position.
3) Press “Enter” key to confirm and exit the saving or recalling menu. If the saving is selected, the values
of List2-1 parameters will be saved in the specified location in EEPROM; if the recalling is selected, the
values of List2-1 parameters should be the values saved in specified location in EEPROM.
Example1: Set the mode to CCL; set the current value to 2A; turn on the load; save the settings in location
0 in EEPROM; the next time the load is turned on, these settings will become the power-on settings.
Procedures /// Operation Descriptions /// Display
1 /// Set the mode to CCL, set the current value to 2A; turn on the load (please refer to 4.5.1 section for
detailed operations). ///
2 /// Press “Shift”+“1Store” keys to enter into the saving menu. /// SAVE 0
3 /// Press Entry keys or the knob to select the location 0 for saving. /// SAVE 0
4 /// Press “Enter” key to confirm. ///