5.3. Connecting to the Power Supply
Connect the positive pole of the power supply to the INPUT + terminal, and connect the negative pole
of the power supply to the INPUT - terminal. If the input is connected reversely, the RV protection sta-
tus of the load will become effective, and please try to connect the power supply to the load again at this time.
After the power supply is correctly connected to the load, long-press “Cancel key to clear the RV
protection status (the relevant details will be described in “Clearing Protection Settings”).
5.4. Turning the Input On/Off
Press “On/Off” key to turn on or turn off the input.
5.5. Basic Tests
5.5.0.
The operating procedures for basic tests are shown as below:
1) Press “Menu” key to enter into the mode selection and parameter setting menu.
2) Use Entry keys or use the knob together with ”, keys to input set value. Use “Cancel” key to modify
the set value or exit the mode selection and parameter setting menu.
3) Press“Enter” key to confirm and exit the mode selection and parameter setting menu.
4) Press “On/Off” key to turn on the load.
5.5.1. CC Mode
CC Mode includes constant current high range (CCH) and constant current low range (CCL).
Example1: Set the load to CCH and set the current value to 5.12A in CV mode. Turn on the load,
and check the present power value. The operating procedures are shown as below:
Procedures /// Operation Descriptions /// Display
1 /// Press “Menu” key to enter into the selectable menu. /// MENU 0 SYST
2 /// Use Entry keys or use the knob to select “0 System Set”,press “Enter” key to confirm and enter into
System Set. /// CC SET 0 Low
3 /// Use Entry keys or use the knob to select “1 High”,press “Enter” key to confirm. /// CC Set 1 High
4 /// Press “Cancel” key to exit and return to the basic mode. /// CVH 05.000V
5 /// Press “I-set” key to enter into the parameter setting menu. Use Entry keys or use the knob together
with “t” and “u” keys to input the current value to 5.12. /// SCCH 05.120A
6 /// Press “Enter” key to confirm and exit the parameter setting menu. /// CCH 05.120A
7 /// Press “On/Off” key to turn on the load. ///
For the above operating procedures, the corresponding SCPI commands should be:
INPUT Off ; Turns off the load