AXIOMET AX-EL300W60A [31/72] Reading remote programming errors

AXIOMET AX-EL300W [31/72] Reading remote programming errors
CPC Power level /// CPC immediate power level
Transient operation /// Transient test
Transient mode /// Transient mode
Transient Htime /// Time for transient high level
Transient Ltime /// Time for transient low level
Transient Rtime /// Time for transient rising edge
Transient Ftime /// Time for transient falling edge
Trigger Function /// Trigger function selection
Trigger source /// Trigger source
Battery Mode /// Battery discharge operation
Battery mini voltage /// Battery discharge termination voltage
Battery discharge current /// Battery discharge current
Voltage on Latch /// Latch the Von point
Voltage on /// Von point for the load
Voltage off /// Voff point for the load
3.12. Reading Remote Programming Errors
The Err annunciator will be turned on when remote programming errors occur. The error codes are shown
as followings:
- -1xx Command errors
- -2xx Execution errors
- -3xx Device-specific errors
- -4xx Query errors
The remote command (SYSTem:ERRor?) can reads back the error codes and error messages when
it is in remote control.
All errors are saved in one error queue. The errors in this error queue are read in the order in which they
occurred. At most 20 error messages can be saved in the error queue. If the occurred errors exceed 20,
the last error in the error queue will be replaced with -350, “Queue overflow”. The load will not save any
additional error message, unless you clear or read errors from the queue. Once the error message is read,
it will be cleared in the error queue.
3.13. Status Report
The electronic load incorporates a status reporting register. Various status conditions of the load can be
reported by querying the status register. The user can make sure which event has been reported through
setting the enable register (It will be introduced in details in “ElectronicLoad SCPI Programming Guide”.).