AXIOMET AX-EL300W60A [48/72] Sequence list operation

AXIOMET AX-EL300W60A [48/72] Sequence list operation
8 /// Use the knob together with ”, keys to set TimeF to 10ms, press “Enter” key to confirm. /// TIMEF
040.00M
9 /// Press “Cancel” key to exit the transient operation menu. ///
10 /// Press “On/Off” key to turn on the load. ///
5.8. Sequence (List) Operation
5.8.0.
After sequence editing, the operating procedures for enabling and disabling sequence operation are shown
as below:
1) Press “List” key to enter into sequence operation menu.
2) Use Entry keys or the knob to set sequence State to On, and press “Enter” key to confirm.
3) Use Entry keys or the knob together with “↑”, “↓” keys to set list number (No.), and press “Enter” key to
confirm.
4) Press “Cancel” key to exit sequence operation menu.
5) Press “Shift”+“4Start” keys to activate sequence operation.
6) Press “Shift”+“5Stop” keys to stop sequence operation.
The operating procedures for sequence operation are shown as below:
1) Press “List” key to enter into sequence operation menu.
2) Use Entry keys or the knob to set sequence State to On, and press “Enter” key to confirm.
3) Use Entry keys or knob to select list number (No.). Press “Enter” key to confirm and recall the sequency
in EEPROM assigned by the sequence number.
4) Use “↓” key to select list memo (Memo).
5) Use knob and “ ”, “ ” keys to edit memo (max. 10 characters).
6) Use “↓” key to select sequence data (Data:<New/Edit>). Use “ ”, “ ” keys keys or knob to select New or
Edit operation.
7) Select “New” operation, and press “Enter key to enter into data clear screen. Use the knob to select
“Yes” and press “Enter” key to confirm, the data is cleared and meanwhile step edit screen appears. Use
” key or knob to select “No”, press “Enter” key to confirm, in the meantime the load returns to sequence
(list) test menu.
8) Select “Edit” operation, and press “Enter” key to enter into step edit screen.
9) Use “↑”, “↓” keys to select parameters.
10) Use Entry keys or use the knob together with ”, keys to input set value; press “Enter key to
confirm and enter into the next step.
11) If it is needed to modify one step, use “↑”, “↓” keys to select this step. The setting operations for this
step are the same as the operating procedure 9.
12) If it is needed to add one step to the last step, use “↓” key to select the new step. The setting operations
for this new step are the same as the operating procedure 9.
13) If it is needed to insert one step above an edited step, use “↑”, “↓” keys to select this edited step. The