AXIOMET AX-EL300W60A [45/72] Transient operation

AXIOMET AX-EL300W60A [45/72] Transient operation
6 /// Press “Cancel” key to exit the main menu. On the display, “s” disappears in CCH status. /// CVH
0.000V
For the above operating procedures, the corresponding SCPI commands should be:
INPUT:SHORT OFF ; Sets the load to exit short circuit operation
INPUT OFF ; Turns on the load
5.7. Transient Operation
5.7.0.
The operating procedures for transient operation are shown as below:
1) Select one basic test mode, press “Enter” key to confirm and exit the parameter setting menu.
2) Press “Tran” key to enter into transient operation.
3) Use “↑”, “↓” keys to select the parameter.
4) Use Entry keys or use the knob together with ”, keys to set values for LevelL, LevelH, TimeL,
TimeH, TimeR, and TimeF. Use Entry keys or knob to set Mode value and set State to On.
5) Press “Cancel” key to exit the transient operation menu.
6) Press “On/Off” key to turn on the load.
7) If it is the pulsed transient operation or the toggled transient operation, one trigger occurs when pressing
“Shift”+“0Trigger” keys or the signal at the trigger input terminal (TRIG IN) is low level.
5.7.1. Continuous Transient Operation
In continuous operation, the load periodically switches between high/low levels. The relevant parameters
such as low level (LevelL), high level (LevelH), low level time (TimeL), high level time (TimeH), time for
rising edge (TimeR), time for falling edge (TimeF).
Example: Set the load voltage value to periodically switch between 1V and 5V; set the time for ris-
ing edge (TimeR) to 10ms; set high level time (TimeH) to 200ms; set time for falling edge (TimeF) to 20ms;
set low level time (TimeL) to 400ms, and the load is in continuous transient operation. The operating
procedures are shown as below:
Procedures /// Operation Descriptions /// Display
1 /// Press “V-set” key to select CVH mode, press “Cancel” key to exit the parameter setting menu. /// CVH
04.000V
2 /// Press “Tran” key to enter into transient operation. /// STATE 0OFF
3 /// Use Entry keys or use the knob together with “ ”, ” keys to set State value to “1 On”, press “Enter”
key to confirm. /// STATE 1ON
4 /// Use Entry keys or use the knob to set Mode value to “Cont”,press “Enter” key to confirm. /// MODE
0Cont
5 /// Use Entry keys or use the knob together with ”, keys to set LevelL to 1, press “Enter” key to
confirm. /// LEVELL 01.000V