6.10. Pulsed Transient Operation Example
This example selects CV pulsed transient operation, selects the bus as the trigger source, and sets a pulse
width of 1ms.
1) ”INPUT OFF” ; Turns off the load’s input
2) ”MODE CVH” ; Selects the CV mode
3) ”TRIG:FUNC TRAN” ; Selects the transient for trigger function
4) ”TRIG:SOUR BUS” ; Selects the bus as the trigger source
5) ”VOLT:LLEV 0.5” ; Sets the transient low level to 0.5V
6) ”VOLT:HLEV 1” ; Sets the transient high level to 1V
7) ”TRAN:HTIM 0.001” ; Sets transient high level time to 1ms
8) ”TRAN:RTIM 0.000250” ; Sets the time for transient rising edge to 250us
9) ”TRAN:FTIM 0.000250” ; Sets the time for transient falling edge to 250us
10) ”TRAN:MODE PULSE ” ; Selects the pulsed transient operation
11) ”TRAN ON” ; Activates the transient operation
12) ”INPUT ON” ; Turns on the load’s input
13) ”*TRG” ; *TRG command generates a 1ms high-level pulse at the load’s input
7. Auto Test Operation
7.1. Main Menu
Press key to enter into main menu of Auto Test. Rotate knob to select menu. Press “Enter key
to enter into the function option on the cursor. Use “↑ key or ↓” key to turn over the LCD screen.
At this time, press “↓” key to enter into the next level submenu. Press “Cancel” key to exit the menu setting.
Please see the list below for main menu content:
Trigger set
/// Function /// Trigger function: 0 input - input On/Off; 1 Level - recall trigger value; 2 List - enable List; 3
Tran - enable Tran
/// Source /// Trigger source selection: 0 External - external key trigger (Shift+“0Trigger”); 1 Pluse -
TTL trigger on the rear panel; 2 Bus - the trigger source is *TRG command; 3 Hold - use the command
“TRIGger:IMMediate” as trigger source
Auto Test
/// State /// Auto test state on/off
/// End Step /// End step (0-14)
/// Step /// Step selection (0-14)
/// Mode /// Mode setting
/// Level /// Setting value
/// Min /// Parameter limit minimum value