MODE CCH ; Sets the mode
CURR 5.12 ; Sets the current value
INPUT ON ; Turns on the load
Example2: Set the current value to 5.8A in CCH. Turn on the load, and there are two ways to set the
current value.
Solution 1:
Procedures /// Operation Descriptions /// Display
1 /// Press “I-set” key to enter into the parameter setting menu. /// SCCH 05.120A
2 /// Use Entry keys or use the knob together with and keys to input the current value to 5.8.
/// SCCH 05.800A
3 /// Press “Enter” key to confirm and exit the parameter setting menu. /// CCH 05.800A
4 /// Press “On/Off” key to turn on the load. ///
Solution 2:
Procedures /// Operation Descriptions /// Display
1 /// Move the cursor to the hundredths with “ ” key. /// CCH 05.120A
2 /// Rotate the knob to set the hundredths to 0. (The set value will be changed when rotating the knob.
The set value will become effective at input at once when the load is turned on.). /// CCH 05.100A
3 /// Move the cursor to the tenths with “ ” key. /// CCH 05. 100A
4 /// Rotate the knob to set the tenths to 8. (The set value will be changed when rotating the knob. The set
value will become effective at input at once when the load is turned on.). /// CCH 05. 800A
5 /// Press “On/Off” key to turn on the load. ///
For the above operating procedures, the corresponding SCPI commands should be:
MODE CCH ; Sets the mode
CURR 5.8 ; Sets the current value
INPUT ON ; Turns on the load
Note:
CCH /CCL will be shown respectively in the lower right corner of the display when the load is in CCH
status or CCL status.
5.5.2. CV Mode
Example1: Set the load to CCH and set the voltage value to 50V in CVH mode. Turn on the load, and check
the present power value. The operating procedures are shown as below:
Procedures /// Operation Descriptions /// Display
1 /// Press “V-set” key to enter into the parameter setting menu. Use Entry keys or use the knob together
with “ ” and “ ” keys to input the voltage value to 50. /// SCVH 50.000V
2 /// Press “Enter” key to confirm and exit the parameter setting menu. /// CVH 50.000V
3 /// Press “On/Off” key to turn on the load. ///
For the above operating procedures, the corresponding SCPI commands should be:
MODE CVH ; Sets the mode