VOLT 50 ; Sets the voltage value
INPUT ON ; Turns on the load
Example2: Set the voltage value to 60V in CV. Turn on the load, and there are two ways to set the voltage
value.
Solution 1:
Procedures /// Operation Descriptions /// Display
1 /// Press “V-set” key to enter into the parameter setting menu. /// CVH 50.000V
2 /// Use Entry keys or use the knob together with and “ keys to input the voltage value to 60. /// CVH
60.000V
3 /// Press “Enter” key to confirm and exit the parameter setting menu. /// CVH 60.000V
4 /// Press “On/Off” key to turn on the load. ///
Solution 2:
Procedures /// Operation Descriptions /// Display
1 /// Move the cusor to the tenths with “ ” key. /// CVH 50.00V
2 /// Rotate the knob to set the tenths to 6. (The set value will be changed when rotating the knob. The set
value will become effective at input at once when the load is turned on.). /// CVH 60.00V
3 /// Press “On/Off” key to turn on the load. ///
For the above operating procedures, the corresponding SCPI commands should be:
MODE CVH ; Sets the mode
VOLT 60 ; Sets the voltage value
INPUT ON ; Turns on the load
Note:
CVH will be shown in the lower right corner of the display when the load is in CV mode.
5.5.3. CR Mode
CR Mode includes the low level (VL CRL), medium level (VL CRM), high level (VL CRH) of the low voltage
range and the low level (VH CRL), medium level (VH CRM), high level (VH CRH) of the high voltage
range.
Example1: Set the load to VHCRL and set the resisitance value to 1.5Ω. Turn on the load, and
check the present power value. The operating procedures are shown as below:
Procedures /// Operation Descriptions /// Display
1 /// Press “Menu” key to enter into the selectable menu. /// MENU 0 SYST
2 /// Use Entry keys or use the knob to select “0 System Set”,press “Enter” key to confirm and enter into
System Set. /// CC Set 1 High
3 /// Select “Syst CR Set” with “↑” and “↓” keys. /// CR Set 2 VH High
4 /// Use Entry keys or use the knob to select “0 Low”,press “Enter” key to confirm. /// CR Set 0 VH Low
5 /// Press “Cancel” key to exit and return to the basic mode. /// CVH 60.000V
6 /// Press “R-set” key to enter into the parameter setting menu. Use Entry keys or use the knob together