If the load is unable to operate normally, please check the setting of Von Point and Voff Point.
3.9.3. Current Limit in CV Mode
The CV Curr Limit is used to limit the maximum input current in CV mode. If the voltage is still larger
than the set level while the current limit has been reached, the load will switch to the CC mode. Please see
figure 3-13. The input will not be turned off in the CV current limit, which is different from the software
current limit.
The CV Curr Limit can be set in main menu, or via the remote command (CV:CURRent:LIMIt
<NRf+>).
1 - Input voltage
2 - Current Limit
3 - Setting Voltage
4 - Load Current
3.9.4. Current Rise Rate
The Current Rise Rate is used to set the current rise rate in CC mode. It can be set in main menu, or via
the remote command (CURRent:RISE:RATE <NRf+>).
If the current rise rate is 0.1A/us, and the current set level is 20A, then the current rise rate is shown as
below when the input is turned on: